Meditech Electronics

Ventilator Drager Evita 2 Dura

₹ 350000.00  Incl. GST ₹ 296610.17 Excl. GST

Description

Ventilator Drager Evita 2 Dura

Ventilation mode

IPPV, IPPVAssist (CMV, CMVAssist)
SIMV, SIMVASB (SIMV, SIMV/Psupp)
MMV, MMVASB (MMV, MMV/Psupp)
BIPAP1), BIPAP1)ASB, BIPAP1)Assist (PCV+, PCV+/Psupp, PCV+Assist)
CPAP, CPAPASB (CPAP, CPAP/Psupp)
APRV (optional)
ILV (optional)

Enhancements

AutoFlowTM – Automatic adaptation of inspiratory flow in volume controlled modes
ATCTM – Automatic Tube Compensation (optional)
NIV - Mask Ventilation (optional)
Ventilation frequency (f): 0 to 100 /min, 0 to150 /min (Neonatal)
Inspiration time (Tinsp): 0.1 to 10 s

Tidal volume (VT) (BTPS):

0.1 to 2.0 L (Adult)
0.02 to 0.3 L (Pediatric)
0.003 to 0.1 L (Neonatal)

Inspiratory flow:

6 to 120 L/min (Adult)
6 to 30 L/min (Pediatric and Neonatal)
Inspiratory pressure: 0 to 80 mbar (cmH2O)
PEEP / intermittent PEEP: 0 to 35 mbar (cmH2O)
Pressureassist (PASB) (Psupp): 0 to 80 mbar (cmH2O)
Rise time for inspiratory pressure: 0 to 2 s
O2 concentration: 21 to 100 Vol.%
• Multi-sense Trigger Criteria: Internal automatic pressure trigger, Flow, Volume (Flow adjustable 0.3 to 15 L/min)

Measured values displayed

Airway pressure: Peak pressure, plateau pressure, mean pressure, PEEP, min. pressure (0 to 99 mbar/cmH2O)
Minute volume (MV), (BTPS): MV, MVspont, MVleak (0 to 99 L/min)
Tidal volume (VT), (BTPS): Inspired VT, expired VT, VTASB (VTPS) (0 to 3999 mL)
Breathing frequency (f): ftotal, fspon, fmand. (0 to 300/bpm)
O2 concentration: (FiO2) Inspired O2 concentration (15 to 100 Vol.%)

Lung mechanics:

 Resistance ((0.0 to 600 mbar/L/s)(cmH2O/L/s)
• Compliance ((0.0 to 300 mL/mbar)(mL/cmH2O)
Breathing gas temperature: 18°to 51°C
Waveforms: Airway pressure-time, flow-time, volume-time, ...
Trends (optional): FiO2, MV, VT, f, PEEPi, R, C, etCO2, ...
Loops (optional): Paw-V, V-Flow, Flow-Paw, ...
Capnography (etCO2) (optional): 0 to 100 mmHg
CO2 production (VCO2): 0 to 999 mL/min, STPD
Serial dead space Vds: 0 to 999 mL, BTPS
Dead space ventilation (Vds/VT): 0 to 99%

Alarms / Monitoring

Airway pressure: High/Low
Expired minute volume: High/Low
Tidal volume: High
Apnea alarm time: 5 to 60 s
Spontaneous breath frequency: High
Inspired O2 concentration: High/Low
Breathing gas temperature: High
SpO2 pulse (optional): High/Low
etCO2 (optional): High/Low

Operating data

Mains power connection: 100 to 240 V, 50/60 Hz, 10 to 30 V DC (optional)
Power consumption: Approx. 125 W
Gas supply operating pressure: O2, air: 2.7 to 6 bar/39 to 87 PSI

Physical Specifications

Dimensions ventilator (WxHxD): 530x290x450 cm (20.9x11.4x17.7 inches) (without trolley)
Weight basic unit: Approx. 27 kg (60 lbs.)
Diagonal screen size: 6.5" TFT LCD color display

Machine outputs

Digital output: Output and reception via an RS 232 C interface
Digital output (optional): For output and reception via two RS 232 C interfaces
Analog output (optional): For analog output of two measured values

SELLER: Meditech Electronics

Ventilator Drager Evita 2 Dura

Product code (MPN): ITB-C32B3003
Availability:In Stock
₹ 350000.00  Incl. GST ₹ 296610.17 Excl. GST
  • Description
    Ventilator Drager Evita 2 Dura

    Ventilation mode

    IPPV, IPPVAssist (CMV, CMVAssist)
    SIMV, SIMVASB (SIMV, SIMV/Psupp)
    MMV, MMVASB (MMV, MMV/Psupp)
    BIPAP1), BIPAP1)ASB, BIPAP1)Assist (PCV+, PCV+/Psupp, PCV+Assist)
    CPAP, CPAPASB (CPAP, CPAP/Psupp)
    APRV (optional)
    ILV (optional)

    Enhancements

    AutoFlowTM – Automatic adaptation of inspiratory flow in volume controlled modes
    ATCTM – Automatic Tube Compensation (optional)
    NIV - Mask Ventilation (optional)
    Ventilation frequency (f): 0 to 100 /min, 0 to150 /min (Neonatal)
    Inspiration time (Tinsp): 0.1 to 10 s

    Tidal volume (VT) (BTPS):

    0.1 to 2.0 L (Adult)
    0.02 to 0.3 L (Pediatric)
    0.003 to 0.1 L (Neonatal)

    Inspiratory flow:

    6 to 120 L/min (Adult)
    6 to 30 L/min (Pediatric and Neonatal)
    Inspiratory pressure: 0 to 80 mbar (cmH2O)
    PEEP / intermittent PEEP: 0 to 35 mbar (cmH2O)
    Pressureassist (PASB) (Psupp): 0 to 80 mbar (cmH2O)
    Rise time for inspiratory pressure: 0 to 2 s
    O2 concentration: 21 to 100 Vol.%
    • Multi-sense Trigger Criteria: Internal automatic pressure trigger, Flow, Volume (Flow adjustable 0.3 to 15 L/min)

    Measured values displayed

    Airway pressure: Peak pressure, plateau pressure, mean pressure, PEEP, min. pressure (0 to 99 mbar/cmH2O)
    Minute volume (MV), (BTPS): MV, MVspont, MVleak (0 to 99 L/min)
    Tidal volume (VT), (BTPS): Inspired VT, expired VT, VTASB (VTPS) (0 to 3999 mL)
    Breathing frequency (f): ftotal, fspon, fmand. (0 to 300/bpm)
    O2 concentration: (FiO2) Inspired O2 concentration (15 to 100 Vol.%)

    Lung mechanics:

     Resistance ((0.0 to 600 mbar/L/s)(cmH2O/L/s)
    • Compliance ((0.0 to 300 mL/mbar)(mL/cmH2O)
    Breathing gas temperature: 18°to 51°C
    Waveforms: Airway pressure-time, flow-time, volume-time, ...
    Trends (optional): FiO2, MV, VT, f, PEEPi, R, C, etCO2, ...
    Loops (optional): Paw-V, V-Flow, Flow-Paw, ...
    Capnography (etCO2) (optional): 0 to 100 mmHg
    CO2 production (VCO2): 0 to 999 mL/min, STPD
    Serial dead space Vds: 0 to 999 mL, BTPS
    Dead space ventilation (Vds/VT): 0 to 99%

    Alarms / Monitoring

    Airway pressure: High/Low
    Expired minute volume: High/Low
    Tidal volume: High
    Apnea alarm time: 5 to 60 s
    Spontaneous breath frequency: High
    Inspired O2 concentration: High/Low
    Breathing gas temperature: High
    SpO2 pulse (optional): High/Low
    etCO2 (optional): High/Low

    Operating data

    Mains power connection: 100 to 240 V, 50/60 Hz, 10 to 30 V DC (optional)
    Power consumption: Approx. 125 W
    Gas supply operating pressure: O2, air: 2.7 to 6 bar/39 to 87 PSI

    Physical Specifications

    Dimensions ventilator (WxHxD): 530x290x450 cm (20.9x11.4x17.7 inches) (without trolley)
    Weight basic unit: Approx. 27 kg (60 lbs.)
    Diagonal screen size: 6.5" TFT LCD color display

    Machine outputs

    Digital output: Output and reception via an RS 232 C interface
    Digital output (optional): For output and reception via two RS 232 C interfaces
    Analog output (optional): For analog output of two measured values

  • FAQs
    Need some help? Ask us!